How do I insert the height adapter correctly into 3Shape lab scanner?

How do I insert the height adapter correctly into 3Shape lab scanner?

09/11/2023

Optimizing height adapter alignment

A height adapter’s magnets are placed in an equilateral triangle, so it’s possible to insert the adapter into the scanner in 3 ways.

The correct position is when lines on rotation plate and height adapter match up and face the front of the scanner. 

If the height adapter is inserted incorrectly, calibration will not pass, and various errors may come up during scanning (see article “Calibration failed” errors during lab scanner calibration" for more information).

Height adaptor for E-scanners should not be used for F8 and the other way around.

 

 